Stimulation & Recording Products

Use our Front Ends to independently control up to 512 channels of stimulation and recording.


  • A complete Grapevine system includes Front End headstages that connect to a Grapevine processor
  • Signals are digitized in the Front Ends for low-noise neural recording
  • Fast-settle features allow for stimulus artifact rejection and rapid recording recovery following stimulation

Primary Use Cases

  • Stimulation
  • LFP
  • ECoG
  • Single/multi-unit activity

Technical Details

Current-Controlled Stimulation
Resolution 1, 2, 5, 10, or 20 µA/step (programmable)
Stimulation range ± 127 steps (up to ± 1.5 mA)
Compliance voltage ± 8.5 V

Recording, Single Selectable Reference
Resolution16-bit - 0.125, 0.25, or 0.5 μV/bit (programmable)
Input range± 4, 8, or 12 mV (set by resolution)
Sampling frequency30 ksps
Hardware filtering1.0 Hz HPF & 7.5 kHz LPF
Impedance measurement10 k Ω to 400 k Ω electrodes
Input-referred noise< 2.1 μVrms
Recording recorvery, post stimulationless than 1 ms

Recording Products

Gather precise recordings from up to 32 high-impedance microelectrodes.


  • Ideal for both animal and clinical studies, with IRB approval
  • Features high signal fidelity through on-board filtering, amplification, and 16-bit digitization
  • Alternate between two connected reference electrodes via two small switches on the unit
  • Check impedance of electrodes with recording Front Ends in Trellis Software

Primary Use Cases

  • LFP
  • ECoG
  • single/multi-unit activity

Technical Details

Recording, single selectable reference
Resolution 16-bit - 0.125, 0.25, or 0.5 μV/bit
(programmable)
Input Range ± 4, 8, or 12 mV
(set by resolution)
Sampling frequency 30 ksps
Hardware filtering 0.1 - 100 Hz HPF (programmable) 7.5 kHz LPF
Impedance Measurement 10 k Ω to 1.0 MΩ electrodes
Input - referred noise < 2.1 μVrms

Specialty Recording Products

Designed specifically for your muscle and surface recording experiments.


Primary Use Cases

  • EMG
  • EEG

Technical Details

EMG Recording - single selectable reference
Resolution 24-bit (0.022 μV/bit)
Input Range ± 150 mV
Sampling frequency 2 ksps or 7.5 ksps
Hardware filtering 500 Hz LPF
Input - referred noise < 1.0 μVrms


Surf S2 Recording - single selectable reference
Resolution 16-bit - 0.125, 0.25, or 0.5 μV/bit
(programmable)
Input Range ± 4, 8, or 12 mV
(set by resolution)
Sampling frequency 30 ksps
Hardware filtering 0.1-100 Hz HPF (programmable), 7.5 kHz LPF
Impedance Measurement 10 k Ω to 1.0 MΩ electrodes
Input - referred noise < 1.0 μVrms